The LF-200 zirconia oxygen analyzer was developed for automatic gas
filling packaging equipment based on Toray Engineering's long track
record with oxygen analyzers for the food packaging industry.
This analyzer is equipped with our unique, compact zirconia sensor to
enable stable and highly accurate measurement over the long term, and
an oxygen sensor protection mechanism for unprecedented durability. This
is an easy to operate, compact oxygen analyzer which is ideal for
integration into automatic packaging equipment for foods,
pharmaceuticals and other products.
Features- The system is equipped with a new type of filter to improve
measurement of gases with dispersed oils. The filter can be replaced
without removing the analyzer from the panel.
- Suction pump protection function
- Enables pump protection while avoiding problems due to on-site cleaning caused by blowing air.
- Internal pipe flow which minimizes gas stagnation at points of measurement.
- AC200V is provided standard, for ease of integration into packaging equipment (AC100V is optional).
LF-200 Internal Flow Diagram
Equipment Specifications
Type |
Panel-mounted (Portable type is optional) |
Display system |
Digital 4-digit display (O2 concentration) |
Measurement range |
1 ppm-100 vol%O2 |
Display range |
0-1000 ppm / 1 / 10 / 100 %, automatic
*The 1000 ppm range is not guaranteed |
Output range |
0-0.5 / 5 / 10 / 100 %, Fixed |
Sampling system |
Two methods: continuous suction with a built-in pump, and injection
*Injection method is not guaranteed |
Supply gas rate |
1000 mL/min |
Sensor gas rate |
60 ± 10 mL/min |
Gas connection |
INLET (Sample gas inlet): Rc1/8
BYPASS OUT (Bypass outlet): Rc1/4
OUTLET 1 (Sample gas outlet1): Rc1/4
OUTLET 2 (Sample gas outlet2): Rc1/4 |
Reference gas |
Air |
Weight |
Approx. 7 kg |
Paint color |
Front, back: Silver metallic
Cover: Stainless |
CharacteristicsRepeatability |
Within ±1 % of full scale (at or above 0-1 % range)
Within ±2 % of full scale (below 0-1 % range)
(0-1000 ppm is not guaranteed) |
Air point stability |
Within ±0.5 % of full scale/24h |
Measurement value output |
External output: DC 4-20 mA |
Concentration alarm output |
LED display (Upper/Lower limit)
Contact output: 2a individual output
Contact: AC 30 V DC 30 V, 0.5 A |
Self-diagnosis functions |
Warmup error, Thermocouple error
Internal temperature error
CPU error, Furnace temperature error
Asymmetry potential error, Sensor resistance error
AIR point calibration error
SPAN point calibration error |
Sample gas conditions |
Must not contain flammable components, halogen, silica, corrosive components or water droplets.
Pressure: Pressurized at 0.03 MPa max.
Flow rate: 1,000-2,000 mL/min
Temperature: 50 °C max.
Humidity: Dew point lower than ambient temperature |
Power supply |
Voltage: AC 200 V, Wired to terminal block
Frequency: 50/60 Hz |
Set up conditions |
Set up location: Indoors, Non-explosion-proof zone
Ambient temperature: 0-40 °C
Ambient humidity: 45-85 %RH, No condensation |
Equipment options |
Communication output (RS-232C compliant, unidirectional)
Activated carbon filter, Cotton filter
Power supply voltage AC 100 V | Engineering Specifications Related Products |
|